Operator’s Manual 2052 O’Neil Rd, Macedon NY 14502 Telephone: (315) 986-8090 Fax: (315) 986-8091 www.ankom.com Rev C 1/14/15

Strany 2

Operator’s Manual pg. 10 Rev C 1/14/15 Instrument Installation Site Requirements To install and operate the ANKOM200 Fiber Analyzer you wil

Strany 3 - Table of Contents

Operator’s Manual Rev C 1/14/15 pg. 11 Fiber Analysis Support Items The following support items are needed to perform the fiber analy

Strany 4

Operator’s Manual pg. 12 Rev C 1/14/15 This page intentionally left blank

Strany 5 - Operating Environment

Operator’s Manual Rev C 1/14/15 pg. 13 ADF Analysis ADF Calculation ADF contained within a food or feed sample can be calculated usin

Strany 6 - Contact Information

Operator’s Manual pg. 14 Rev C 1/14/15 8. Record the weight of the sample (W2). 9. Seal the filter bag within 4mm of its open end. Keep

Strany 7 - Instrument Description

Operator’s Manual Rev C 1/14/15 pg. 15 ADF Analysis Procedure using the ANKOM200 Fiber Analyzer To perform the ADF analysis on prepar

Strany 8

Operator’s Manual pg. 16 Rev C 1/14/15 17. Repeat steps 12 - 16 to accomplish the second rinse. 18. Make sure that the HHEEAATT button i

Strany 9 - Safety Precautions

Operator’s Manual Rev C 1/14/15 pg. 17 NDF Analysis NDF Calculation NDF contained within a food or feed sample can be calculated usin

Strany 10 - Instrument Installation

Operator’s Manual pg. 18 Rev C 1/14/15 8. Record the weight of the sample (W2). 9. Seal the filter bag within 4mm of its open end. Keep

Strany 11 - Fiber Analyzer

Operator’s Manual Rev C 1/14/15 pg. 19 NDF Analysis Procedure using the ANKOM200 Fiber Analyzer To perform NDF analysis on prepared s

Strany 12

This page intentionally left blank

Strany 13 - ADF Analysis

Operator’s Manual pg. 20 Rev C 1/14/15 18. After the third rinse, open the Vessel Lid and remove the Bag Suspender from the Vessel. 19. R

Strany 14

Operator’s Manual Rev C 1/14/15 pg. 21 Crude Fiber Analysis Crude Fiber Calculation Crude Fiber contained within a food or feed sampl

Strany 15

Operator’s Manual pg. 22 Rev C 1/14/15 5. Place an empty filter bag in the Bag Holder in an open position. 6. Tare the weight of the e

Strany 16

Operator’s Manual Rev C 1/14/15 pg. 23 Crude Fiber Analysis Procedure using the ANKOM200 Fiber Analyzer To perform Crude Fiber analys

Strany 17 - NDF Analysis

Operator’s Manual pg. 24 Rev C 1/14/15 16. When the Timer sounds, the rinse is complete. Slowly open the Exhaust Valve to drain the hot wa

Strany 18

Operator’s Manual Rev C 1/14/15 pg. 25 37. Remove the samples from the oven and place them directly in a Desiccant Pouch.

Strany 19

Operator’s Manual pg. 26 Rev C 1/14/15 This page intentionally left blank

Strany 20

Operator’s Manual Rev C 1/14/15 pg. 27 Periodic Maintenance Initial Maintenance (during the first 20 hours of operation) Follow the

Strany 21 - Crude Fiber Analysis

Operator’s Manual pg. 28 Rev C 1/14/15 Replacing the Fuse in a 120V instrument To replace the fuse in a 120V ANKOM200 Fiber Analyzer, follo

Strany 22

Operator’s Manual Rev C 1/14/15 pg. 29 Check the Agitation System The agitation system should be checked every three to six months

Strany 23

Table of Contents Introduction ...

Strany 24

Operator’s Manual pg. 30 Rev C 1/14/15 Check the Bag Suspender The Bag Suspender should be checked every three to six months or

Strany 25

Operator’s Manual Rev C 1/14/15 pg. 31 Appendix A – ADF Method Acid Detergent Fiber in Feeds - Filter Bag Technique (for A200 and A20

Strany 26

Operator’s Manual pg. 32 Rev C 1/14/15 Calculations % ADF (as-received basis) = 100 x (W3 – (W1 x C1)) W2 Where: W1 = Bag tare weight

Strany 27 - Periodic Maintenance

Operator’s Manual Rev C 1/14/15 pg. 33 Appendix B – NDF Method Neutral Detergent Fiber in Feeds - Filter Bag Technique (for A200 and

Strany 28

Operator’s Manual pg. 34 Rev C 1/14/15 Calculations % NDF (as-received basis) = 100 x (W3 – (W1 x C1)) W2 Where: W1 = Bag tare weight

Strany 29

Operator’s Manual Rev C 1/14/15 pg. 35 Appendix C – Crude Fiber Method (AOCS Ba 6a-05) Crude Fiber Analysis in Feeds - Filter Bag Tec

Strany 30 - Storage

Operator’s Manual pg. 36 Rev C 1/14/15 Calculations % Crude Fiber = 100 x (W3 – (W1 x C1)) W2 Where: W1 = Bag tare weight W2 = Sample

Strany 31 - Appendix A – ADF Method

Operator’s Manual Rev C 1/14/15 pg. 37 Collaborative Laboratory No.Rep Whole CornCattle FeedAlfalfaWhole SoyPoultry StarterCalf

Strany 32

Operator’s Manual pg. 38 Rev C 1/14/15 This page intentionally left blank

Strany 33 - Appendix B – NDF Method

Operator’s Manual Rev C 1/14/15 pg. 39 Appendix D – Wiring Diagram

Strany 34

This page intentionally left blank

Strany 35

Automation saves time and money! ANKOM Technology is an international company with products that include… TDF Dietary Fiber Analyzer  Autom

Strany 36

Operator’s Manual Rev C 1/14/15 pg. 5 Introduction ANKOM Technology designs, manufactures, and markets instruments and suppor

Strany 37 - Rev C 1/14/15 pg. 37

Operator’s Manual pg. 6 Rev C 1/14/15 Contact Information At ANKOM Technology we are committed to your total satisfaction and therefore alw

Strany 38

Operator’s Manual Rev C 1/14/15 pg. 7 Instrument Description General Description The ANKOM200 Fiber Analyzer is designed to efficie

Strany 39 - Appendix D – Wiring Diagram

Operator’s Manual pg. 8 Rev C 1/14/15 Internal Components (Rear View) External Components (Rear View) Maintenance Alert Collector (part #

Strany 40

Operator’s Manual Rev C 1/14/15 pg. 9 Safety Precautions Hazardous Pressure – The contents of the Vessel are hot and unde